88 pro+ Surface Instruction for Use (209196 REV 8)
Cleaning and Disinfecting
Caution—Bleach may result in damage to the surface. Bleach can
only be used for 6 applications for the life of the top cover
(2 years). For more information, see “Expected Life” on page 95.
Remove any disinfectant residue prior to and after the use of
bleach with a new or clean cloth/wipe soaked in tap water.
Not all cleaners and disinfectants listed in Table 1 may be approved
for sale in your country. Always refer to the local regulations for
applicable approval, listed in Table 1, cleaners and disinfectants. For
questions, contact your Baxter representative.
When you perform the detailed cleaning steps, please note the following:
• A microfiber cloth or ready-to use wipe is recommended as the
wiping cloth
.
• Always replace the wiping cloth when visibly soiled.
• Always replace the wiping cloth between steps (spot clean, clean,
and disinfect).
• Always use Personal Protective Equipment (PPE).
• Adjust the bed position, siderails, headboard, and footboard as
needed for ease of cleaning and disinfection.
Prepare the Bed and Surface for Cleaning and Disinfecting
a. If applicable, fully extend the foot section.
b. Unplug the bed and surface (if applicable).
Warning—Store the status indicator during cleaning and
disinfecting of the surface. Failure to do so could cause injury or
equipment damage
.
c. If applicable, put the MCM status
indicator in the storage location on
either side of the surface.