pro+ Surface Instruction for Use (209196 REV 8) 85
Cleaning and Disinfecting
CLEANING AND DISINFECTING
The Cleaning and Disinfecting section of this manual only addresses the
pro+ surface. Refer to the appropriate bed user manual for instructions on
how to clean and disinfect the bed.
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—The surface must be cleaned and disinfected per the
cleaning and disinfecting instructions.
• Warning—The potential for electrical shock exists with electrical
equipment. Failure to follow facility protocol could cause death
or serious injury.
• Warning—Do not reuse wiping material for multiple steps or on
multiple products.
• Warning—Harmful cleaning solutions may cause skin rash
and/or irritation upon contact. Follow the manufacturer’s
instructions found on the product label and Safety Data Sheet
(SDS).
• Warning—Lift and move items correctly. Do not twist, and seek
assistance when necessary. Make sure the bed is at a correct
height to lift items off the bed.
• Warning—Fluid spills on to the surface electronics could cause a
hazard. If such a spill occurs, unplug the bed and surface and
remove them from service. When fluid spills occur outside of
what is seen in normal use, immediately do as follows:
a. Unplug the bed and surface from its power source.
b. Remove the patient from the bed.
c. Clean the fluid spill from the bed system.
d. Have maintenance examine the system completely.
e. Do not put the surface back into service until it is completely
dry, tested, and found to be safe to operate.