pro+ Surface Instruction for Use (209196 REV 8) 39
VERSACARE Bed—pro+ Surface Installation Instructions
10. Use the surface handles on the side of the
surface to lift and install the surface on the
bed. Make sure the logo side is up and at
the foot end of the bed.
11. Put the bed in this configuration:
• Knee in the highest position
•Head in the highest position
•Bed in the highest position
Step 2: Route the Long Surface Power Cord
(P3255A01, P3255A02, and P3255A05)
Warning—Failure to route the surface power cord correctly could
cause a tripping hazard and/or the cord to be pinched during bed
functions. Injury and/or equipment damage could occur.
The surface power cord is
routed along the patient-left
side of the VersaCare Bed.
NOTES:
•Do not use cable ties to attach the power cord to the bed.
• A label on the surface power
cord shows the routing and
the attachment locations for
the surface power cord on the
bed frame.