Cleaning and disinfection
78
7990044_030_10 – 2078511 – 2023-01-19
16. Clean the hook couplings at the head end of the operating
table to remove all residues.
Move the adjustable hook couplings all the way up and clean
them to remove all residues. Then move the hook couplings
all the way down and clean again so there are no residues.
17. Extend the operating tabletop with the [i16] key all the way to
the head end.
18. Clean the extended operating tabletop from underneath so
there are no residues.
19. Extend the operating tabletop with the [i17] key all the way to
the foot end.
20. Clean the extended operating tabletop from underneath so
there are no residues.
21. Move the operating tabletop with the [i5] key all the way up.