BD BACTEC™ FX40 Instrument User’s Manual
142
09
Instrument X: One
or more stations
has had a
measurement
failure or has had a
vial presence
switch failure.
Open the FX40 to
resolve error(s).
Instrument has detected a
condition that could represent
a measurement failure or
partially seated vial.
Alert clears when the instrument
detects that the failure no longer
exists.
Make sure all vials are fully seated in
the stations.
If alert does not clear, block the station
and contact BD.
Stations are marked as Unusable.
Refer to Section 7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
13
Database
Reinitialized.
Set during startup when the
instrument detects corruption
in the database.
Alert clears when the instrument
reinitializes the database and
eliminates the corruption.
14
Time offset
(SESC) will inhibit
the movement of
vials between
instruments.
Contact BD.
A time mismatch has
occurred between multiple
instruments.
This error is set and cleared
automatically. Contact your local BD
representative if error persists/ recurs.
15
Contains a vial
which may have a
reading gap due to
an invalid time
offset (SESC).
Consult manual.
Most likely cause is that a vial
was moved between
instruments with mismatched
time offsets.
Any vials with reading gaps are
marked as Affected Vials. Refer to
Section 7.3.3 for instructions on
Affected Vials.
16
BD EpiCenter
Communications
failure.
This alert is detected when
BD EpiCenter™ is configured
and cannot be reached.
Alert clears itself when communication
with BD EpiCenter™ is reestablished.
17
LIS server not
responding to
uploads.
This alert is detected when
LIS is configured and cannot
be reached.
Alert clears itself when communication
with LIS is reestablished.
18
LIS interface
offline.
Alert is set when LIS library
returns any of errors below to
the Fx application.
LIS_SYSTEM_ERROR:
UNSUPPORTED_ CONFIG:
LIS_ASSERT_ERROR:
DEBUG_PROBLEM:
Communications problem between
instrument and LIS system. Refer to
the BD LIS Interface specification.
Error
No.
Message Possible Cause(s) Corrective Action(s)