BD BACTEC™ FX40 Instrument User’s Manual
144
31
Station X-X0:
Reading Collection
Database Object
Reinitialized.
An individual reading or a
reading collection is
corrupted.
A sector on the flash drive is corrupted
or a bad checksum is encountered on
a reading object or collection. One or
more readings have been lost.
Message is informational. If four
consecutive readings become
corrupted, then a reading gap will
occur and the vial will automatically
become affected.
32
Barcode Reader n:
Cannot determine
type.
Instrument cannot
communicate with barcode
reader to determine the
barcode reader’s type.
Communication attempt continues
every two minutes until successful
communication with the barcode
reader is established.
33
Program download
failure.
Set when a microprocessor
download fails to complete
successfully.
Reboot instrument.
34
Watchdog Timeout
Failure, Notify
Becton Dickinson.
An error occurred in the tablet
PC software.
Reboot the tablet using the
Maintenance > Utilities > Reboot
function.
36
Reboot Reason. Message sent to System Alert
display only (Info detail
window). Instrument has
rebooted for one of the
following causes:
1. Unknown Cause
2. Software Upgrade Initiated
3. Software Upgrade
Completed
4. Power fail
5. Database Reinitialized
6. Software Assert
7. Watchdog Timeout
8. Startup Config Screen
Reboot
9. Invalid WD Count
10. Syscall Fault
11. Pure Virtual Call in CRT
12. Structured exception
handling event error
13. Program termination
called by CRT
14. Invalid Reason Code
15. Stack Fault
16. Allocation Fault
17. Reboot button pressed
18. Application is restarting to
complete software upgrade
If error recurs, contact BD.
For Reasons 4 and 5, if power is lost
for more than 40 minutes, all vials in
the instrument are marked as Affected
Vials. Refer to Section 7.3.3 for
instructions on Affected Vials.