EasyManua.ls Logo

Biotek PowerWave - Page 127

Biotek PowerWave
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PowerWave Operator’s Manual 4-11
C
A
L
I B
R
A
T
I O
N
F
I L
T
E
R
:
4
0
5
4
0
5
4
5
0
4
9
0
6
3
0
Figure 4-8: Selecting the Filter Wavelength during Calibration Setup
Using the insert packaged with the calibration plate, press the soft
key beneath the desired filter wavelength to select it.
Then press the ENTER key to advance to the menu that allows
entry of Well Location and calibration values.
W
A
V
E
L
E
N
G
T
H
: 4
0
5
W
E
L
L
: C
0
1
C
A
L
I B
R
A
T
I O
N
V
A
L
U
E
S
: 0
. 0
0
0
Figure 4-9: Entering Calibration Values using the Cal Plate Insert.
After each entry, press ENTER to advance to the next consecutive
well location. Enter the values listed on the Calibration insert
packaged with the plate. The cursor will be positioned under the
first editable field.
When all values have been entered, press the Main Menu key.

Table of Contents

Related product manuals