4-12 Performance Verification
Running the Calibration Plate Test
Power up the PowerWave. From the reader’s Main Menu display:
• Press the soft key beneath UTIL to access the Utility Options menu.
• At the SELECT UTILITY OPTION menu (Figure 4-4), press the soft
key beneath TESTS.
• At the SELECT TEST menu, press soft key #4 to select CALPLATE.
S
E
L
E
C
T
T
E
S
T
?
S
Y
S
T
E
M
C
H
K
S
U
M
C
A
L
P
L
T
E
Figure 4-10: Selecting the Calibration Plate Test
• Select the appropriate wavelength at the CALIBRATION FILTER screen
by pressing the soft key beneath the desired value. Press ENTER to save the
value and advance to the next screen.
C
A
L
I B
R
A
T
I O
N
L
A
M
B
D
A
: 4
0
5
4
0
5
4
5
0
4
9
0
6
3
0
Figure 4-11: The Calibration Filter Selection screen
• When prompted, insert the calibration plate into the PowerWave’s
plate carrier, and press the READ key to begin the calibration
program.
The Calibration Plate Analysis Report (page 4-10) will be sent to a printer
when the test is run.