4-12 Performance Verification
Running the Calibration Plate Test
Power up the PowerWave. From the reader’s Main Menu display:
Press the soft key beneath UTIL to access the Utility Options menu.
At the SELECT UTILITY OPTION menu (Figure 4-4), press the soft
key beneath TESTS.
At the SELECT TEST menu, press soft key #4 to select CALPLATE.
S
E
L
E
C
T
T
E
S
T
?
S
Y
S
T
E
M
C
H
K
S
U
M
C
A
L
P
L
T
E
Figure 4-10: Selecting the Calibration Plate Test
Select the appropriate wavelength at the CALIBRATION FILTER screen
by pressing the soft key beneath the desired value. Press ENTER to save the
value and advance to the next screen.
C
A
L
I B
R
A
T
I O
N
L
A
M
B
D
A
: 4
0
5
4
0
5
4
5
0
4
9
0
6
3
0
Figure 4-11: The Calibration Filter Selection screen
When prompted, insert the calibration plate into the PowerWave’s
plate carrier, and press the READ key to begin the calibration
program.
The Calibration Plate Analysis Report (page 4-10) will be sent to a printer
when the test is run.