A-2 Decontamination
Procedure
A solution of .5% Sodium Hypo Chlorite (NaOCL) solution (Bleach)
is used. Commercial bleach is 5% NaOCL; household bleach is 3%
NaOCL. When using commercial bleach use a 10:1 mixture; if using
household bleach, a 6:1 mixture is required. This is a caustic solution.
It is important to wear gloves and eye protection when handling it.
Wipe down the carrier and all exposed surfaces of the unit with the
bleach solution. Remove the top shroud of the instrument and wipe
down the top surface of the instrument base, as well as the inside of
the top shroud.
Reassemble the unit and discard the used gloves and towels.