3-24 Operation
P-ACROSS enables a blank in every even-numbered column to be
subtracted from the well to the left of it in every odd column. Use
manual mapping to set up the appropriate map by placing the
standards, controls and samples in only the odd columns.
P-DOWN enables a blank in the B, D, F and H rows to be subtracted
from the well above in the A, C, E and G rows. Use manual
mapping to set up the appropriate map by placing the standards,
controls, and samples in only the A, C, E, and G rows.
CONSTANT BLANK VALUE
This entry screen (Figure 3-28) only appears when a Constant Blank map is
selected. The user enters a value that is subtracted from each well on the
plate.
E
N
T
E
R
B
L
A
N
K
I N
G
C
O
N
S
T
A
N
T
: 1
: 2
0
Figure 3-28: Constant Blank Value screen
Use the NUMERIC keys to enter the value. The range is 0.000 to
3.000. The cursor is positioned at the first editable field and
advances automatically.
Use the CLEAR key to clear the value on the display.