PowerWave Operator’s Manual 3-15
NUMBER OF KINETIC POINTS SELECTION
Use this screen (Figure 3-19) to select the number of sequential kinetic points
to calculate the steepest Rate, or the R squared at the steepest Rate.
K
I N
E
T
C
P
O
I N
T
S
: 3
A
L
L
P
O
I
N
T
S
Figure 3-19: Kinetic Points Used entry screen
• Use the NUMERIC keys to input the number of points. The range is
2 to MAX where max is the total number of reads.
• Use SOFT KEY 1 or 2 to select All points.
iii Note: If the number of kinetic reads is changed during assay definition, be
certain also to adjust the kinetic points. The number of kinetic points must be
equal to or less than the total kinetic reads.
ONSET OD SELECTION
Use this screen (Figure 3-20) to enter the onset OD value.
E
N
T
E
R
O
N
S
E
T
O
D
:
1
. 2
3
4
Figure 3-20: Onset OD screen
• Use NUMERIC keys to enter the onset OD. 3.000 Abs is the
maximum value.
• Use the LEFT and RIGHT ARROW keys to move the cursor within
the entered OD field.