owerWave Operator’s Manu
2-13
Bulb Removal
• Always wear safety glasses when handling the bulb.
• Turn the PowerWave off and let it sit for 5 minutes before proceeding.
• Remove the bottom plate mounting hardware (6).
• Remove the bulb socket mounting hardware (2) and carefully remove the
bulb housing.
• Remove the old bulb from the socket.
• Using lens paper or gloves, replace the bulb and mount it back into the
bulb housing. There should be no fingerprints on the bulb.
• Replace the bottom using the mounting hardware.