P
owerWave Operator’s Manu
al
2-13
Bulb Removal
Always wear safety glasses when handling the bulb.
Turn the PowerWave off and let it sit for 5 minutes before proceeding.
Remove the bottom plate mounting hardware (6).
Remove the bulb socket mounting hardware (2) and carefully remove the
bulb housing.
Remove the old bulb from the socket.
Using lens paper or gloves, replace the bulb and mount it back into the
bulb housing. There should be no fingerprints on the bulb.
Replace the bottom using the mounting hardware.