2-8 Installation
Setting Up the Serial Port for Communications with Other Devices
The PowerWave has a 25-pin serial (RS232) port located on the rear panel of
the instrument (see Figure 2-7). The serial port allows the reader to
communicate with a computer, using standard communications software
and/or RS232 protocols. Appendix B contains information on required
protocols for computer control of the reader.
Attaching the Serial Cable
Power down the computer and the reader.
Connect the appropriate serial cable to both machines. The serial
port on the reader is a DTE configuration with a 25-pin (pin-
male) D-sub connector. The connector’s pinout is illustrated in
Table 2-2.
Power up the reader and the computer.
Ensure that the PowerWave and the computer are operating with
the same communications settings.

Table of Contents