C-4 Error Codes
Displayed Error
ERROR 0601 LAMBDA #1 Gain out of range
ERROR 0602 LAMBDA #2 Gain out of range
ERROR 0603 LAMBDA #3 Gain out of range
ERROR 0604 LAMBDA #4 Gain out of range
ERROR 0605 LAMBDA #5 Gain out of range
ERROR 0606 LAMBDA #6 Gain out of range
Errors 601-606 indicate that the gain for a specific wavelength is out of the
range necessary to ensure the filter’s performance to specifications.
PROBABLE CAUSE :
• A bad lamp could provide enough reduction in signal to cause
this error. Misaligned optics could also cause this error.
Displayed Error Potential Cause
ERROR 0700 Reader failed noise test
Error 700 indicates that the measurement electronics do not read
repeatably at a maximum gain, with the light blocked.
PROBABLE CAUSE:
• External signals penetrating the measurement circuit. The
bottom and top shrouds, and the filter wheel cover should be
correctly installed. This problem in a correctly assembled unit
could indicate a problems with the reader’s printed circuit board.