C-6 Error Codes
PROBABLE CAUSE:
The measurement electronics full signal level has changed since the
power-up self-check was last run.
Displayed Error
Potential Cause
ERROR 0C00 Printer timeout error
Error C00 indicates that the time allotted for the instrument to make a
valid connection to a printer has expired.
PROBABLE CAUSE:
Printer not connected or powered up. Printer’s parallel port may not
be correctly selected.
Displayed Error
Potential Cause
ERROR 0D00 Reader failed calibration
checksum test
ERROR 0E00 Wavelength not detected in
reader’s wavelength table (1:
MEAS; 2: Ref; 3: sec.
ERROR 0F00 Reader’s wavelength or channel
signal is out of specified range.
ERROR 1000 Required reader configuration
data missing.
ERROR 1100 Failed configuration checksum
test
ERROR 1200 Calibration data missing
ERROR 1300 Motor not correctly homed
ERROR 1400 Assay incubation Error. Assay
requires incubation but the
instrument does not appear to
have an incubator.