EasyManua.ls Logo

Biotek PowerWave - Page 30

Biotek PowerWave
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-6 Introduction
Repeatability (STD): +/- 1% +/- 0.005 Abs from 0 to 2.000 Abs
+/- 3% +/- 0.005 from 2.000 to 2.5 Abs
(measured @ 405 nm)
Display: 2 X 24 Character LCD
Light Source: Xenon Arc Flash bulb
Dimensions: 16.0” deep X 15.5" wide X 9.75" high
40.1 mm deep X 39.3 mm wide X 24.8 mm high
Weight: 40 lb. maximum (18.1 kg)
Environment: Operational temperature 18-40° C
Humidity: 10-85 % non-condensing
Power Consumption: 100VA
Four Voltage Ranges accommodated by voltage selection switch:
Range 1 100 VAC 90-110 VAC, 50-60 Hz
Range 2 120VAC 103-127 VAC, 50-60 Hz
Range 3 230 VAC 207-253 VAC, 50-60 Hz
Range 4 240VAC 216-264 VAC, 50-60 Hz
Incubation (optional):
Temperature Control: 4° over ambient to 50° C
Temperature Variation: +/- 0.5° C @ 37°C ( with the plate sealed).
Optional internal Bar Code reader :
If enabled, the internal bar code reader recognizes a number of common
bar code types during the plate read operation. Specific information about
each bar code type is available from Bio-Tek Technical Services. The
reader’s bar code option is compatible with the following bar code types:
CODABAR UPC
CODE 39 EAN
INTERLEAVED 2 of 5 MSI
CODE 11 PLESSEY
CODE 93 CODE 128

Table of Contents