Setup
18
Rohde and Schwarz FSV/FSW
Note: Refer to the FSV or FSW Operating Manual for more detailed instructions on signal analyzer opera-
tion. The following represents the minimum steps required to perform a recording with the IQC5000.
Setting the IP Address
To set the IP address of the FSV/FSW:
1. Press Setup on the front panel.
2. Press Network + Remote.
3. In the IP Address field, enter the desired IP address, or alterna
tively select Open Dialog “Network Configuration”
to bring up the Microsoft Network Configuration menu.
4. Set the IP address as outlined in "Network Setup" on page 15.
Configure Windows Firewall on the FSV/FSW to Allow DigIConf Traffic
The Windows Firewall needs to be configured on the FSV/FSW to allow DigIConf traffic from the Computer
Workstation to the FSV/FSW. To configure the firewall:
1. Navigate to Control Panel > Windows Firewall > Advanced Setting
s > Inbound Rules.
2. Select New Rule.
3. Select Port and click Next.
4. Select to apply this rule to TCP traffic on specific local port 50
26 and click Next.
5. Select Allow the Connection
and click Next.
6. Select to apply this rule to Domain, Public, and Private interf
aces and click Next.
7. Name the rule DigIConf and click Finish.
(Optional) Running DigIConf on Another Machine
DigIConf software by default is set to run on the FSV/FSW, so that the IQC5000 system can interface with the EX-IQ
(this default setup is highly recommended). If desired, DigIConf may be installed on a workstation, and the EX-IQ
may be attached to and controlled by that workstation. The following instructions detail how to configure DigIConf
on a workstation.
Installing DigIConf
To install the DigIConf software, run the installer included on the EX
-IQ software disc. You will be prompted to
accept the End-User License Agreement, as well as to accept the installation of various drivers. The DigIConf
software must be version 03.01.009.30 or higher.
Running DigIConf
Ensure the hardware is connected according to the steps outlined
in "Hardware Setup – Record" on page 8 with the
exception of the EX-IQ connected via USB to the desired workstat
ion. Once the hardware setup is complete:
1. Start the DigIConf software.
2. Start IQC Control and add a remote device for the FSV/FSW.
3. Set the IP address for DigIConf wit
hin the remote device menu to the IP address of the workstation. Use the
loopback address of 127.0.0.1 if running locally. (See "Connecting/Disconnecting to a Remote Device" on
page 50 and "Managing Remote Devices" on page 51 for more details.)
The Setup Diagram shown in the program menu should reflect the h
ardware setup: