Why is the Manual Vent Valve of BTG Visual-ICE Laboratory Equipment leaking gas?
- AafernandezSep 1, 2025
The Manual Vent Valve may be open. Fully close the Manual Vent Valve.
Why is the Manual Vent Valve of BTG Visual-ICE Laboratory Equipment leaking gas?
The Manual Vent Valve may be open. Fully close the Manual Vent Valve.
What to do if the pressure displayed on the Gas Indicator of BTG Visual-ICE indicates the gas pressure is too low?
Verify the Argon Shutoff Valve is open. Verify that the gas cylinder valve is sufficiently opened to allow gas flow; if necessary, open the valve approximately another half turn. Verify that the cylinder contains sufficient pressure using the gauge on the cylinder. Replace the cylinder, if needed.
What to do if locking bar on the Needle Connection Panel of BTG Visual-ICE cannot be positioned in the LOCKED position?
Verify that all needles in the channel are fully inserted in the needle connection ports. If the Locking Bar may be defective, move the needle to a different channel. Check the Gas Indicator display and verify the system is not pressurized; if there is pressure, end the procedure and vent the system using the automatic vent function.
Why needle cannot be firmly connected in the needle port of BTG Laboratory Equipment?
Ensure that the locking bar is in the UNLOCKED position. If the needle connector may be defective, use a different needle. If residual gas pressure may be in the needle port, use a different channel. Check the Gas Indicator display and if there is pressure in the system, end the procedure and vent the system using the automatic vent function.
Why no helium gas is delivered to the needle in BTG Visual-ICE?
Check that the helium gas cylinder valve is open to allow sufficient gas flow and verify that appropriate pressure is displayed on the Gas Indicator. If the needle may be clogged, press the Stop button to stop all activity on the channel, disconnect the needle, move it to a different channel, and run the test again. If the problem persists, replace the needle with a new needle.
How to fix unresponsive touch screen on BTG Visual-ICE?
Try using the track pad to control the system. If the touch screen may be out of calibration, navigate to the Configure Settings screen using the track pad and re-calibrate the touch screen. You also can turn OFF and restart the system using the power control knob on the front of the system.
What to do if needle is not freezing during the Needle Integrity and Functionality Test of BTG Visual-ICE Laboratory Equipment?
Ensure the argon gas cylinder valve is open to allow sufficient gas flow and verify appropriate pressure is displayed on the Gas Indicator. If the needle may be clogged, try to test it again. If the needle still does not freeze, press the Stop button, disconnect the needle, move it to a different channel, and run the test again. If the problem persists, replace the needle with a new one.
What to do if channel or needle is not recognized by BTG Laboratory Equipment?
Check that the corresponding Channel Locking Bar is fully in the locked position. Ensure at least one needle is inserted into a channel for the channel to be available. If using a needle with a memory chip, try a new needle. If the channel may be defective, do not use it; move the needle(s) to a different channel and conduct the Needle Integrity and Functionality Test again.
What to do if BTG Laboratory Equipment doesn't recognize USB flash drive?
Ensure the USB flash drive is connected to the port with the USB port icon. Try removing the USB flash drive, waiting a few seconds, and reconnecting it. If the problem persists, try using a different Galil Medical supplied USB flash drive or replace the USB flash drive with a new one.
Outlines the medical fields and tissue types the system is designed for.
Provides critical safety warnings and precautions for operating the system.
Describes the primary interface for controlling and monitoring procedures.
Details the panel where cryoablation needles are connected.
Describes the initial screen for user authentication and system access.
The main operational screen for controlling cryoablation procedures.
Outlines the steps for setting up the system before a procedure.
Details the step-by-step process of conducting a cryoablation treatment.
Provides instructions for safely powering down and preparing the system for storage.
Allows customization of system parameters, user accounts, and units.
Details procedures for cleaning the system components and stylus.
Outlines requirements for system servicing and preventive maintenance.
Explains how to restore the system software to a previous version.
Addresses common problems with system components and user errors.
Covers troubleshooting steps for problems related to gas supply and connections.
Specifies the recommended temperature and humidity for system operation.
Lists physical dimensions, weight, and capacity limits of the system.
Provides contact information for Galil Medical Inc.