Questions and Answers

  • S
    Sharon AyersSep 1, 2025
    Why is the Manual Vent Valve of BTG Visual-ICE Laboratory Equipment leaking gas?
    • A
      afernandezSep 1, 2025
      The Manual Vent Valve may be open. Fully close the Manual Vent Valve.
  • D
    Donald MirandaAug 31, 2025
    What to do if the pressure displayed on the Gas Indicator of BTG Visual-ICE indicates the gas pressure is too low?
    • J
      Julia RamseyAug 31, 2025
      Verify the Argon Shutoff Valve is open. Verify that the gas cylinder valve is sufficiently opened to allow gas flow; if necessary, open the valve approximately another half turn. Verify that the cylinder contains sufficient pressure using the gauge on the cylinder. Replace the cylinder, if needed.
  • M
    Morgan GonzalezAug 30, 2025
    What to do if locking bar on the Needle Connection Panel of BTG Visual-ICE cannot be positioned in the LOCKED position?
    • W
      William BullockAug 30, 2025
      Verify that all needles in the channel are fully inserted in the needle connection ports. If the Locking Bar may be defective, move the needle to a different channel. Check the Gas Indicator display and verify the system is not pressurized; if there is pressure, end the procedure and vent the system using the automatic vent function.
  • J
    James HowardAug 29, 2025
    Why needle cannot be firmly connected in the needle port of BTG Laboratory Equipment?
    • C
      Crystal RossAug 29, 2025
      Ensure that the locking bar is in the UNLOCKED position. If the needle connector may be defective, use a different needle. If residual gas pressure may be in the needle port, use a different channel. Check the Gas Indicator display and if there is pressure in the system, end the procedure and vent the system using the automatic vent function.
  • J
    jeremymartinAug 28, 2025
    Why no helium gas is delivered to the needle in BTG Visual-ICE?
    • A
      Angela GrahamAug 28, 2025
      Check that the helium gas cylinder valve is open to allow sufficient gas flow and verify that appropriate pressure is displayed on the Gas Indicator. If the needle may be clogged, press the Stop button to stop all activity on the channel, disconnect the needle, move it to a different channel, and run the test again. If the problem persists, replace the needle with a new needle.
  • S
    Sarah MartinezAug 27, 2025
    How to fix unresponsive touch screen on BTG Visual-ICE?
    • E
      Erin LewisAug 27, 2025
      Try using the track pad to control the system. If the touch screen may be out of calibration, navigate to the Configure Settings screen using the track pad and re-calibrate the touch screen. You also can turn OFF and restart the system using the power control knob on the front of the system.
  • A
    Adam EdwardsAug 26, 2025
    What to do if needle is not freezing during the Needle Integrity and Functionality Test of BTG Visual-ICE Laboratory Equipment?
    • E
      edward93Aug 26, 2025
      Ensure the argon gas cylinder valve is open to allow sufficient gas flow and verify appropriate pressure is displayed on the Gas Indicator. If the needle may be clogged, try to test it again. If the needle still does not freeze, press the Stop button, disconnect the needle, move it to a different channel, and run the test again. If the problem persists, replace the needle with a new one.
  • J
    Jake AndersenAug 25, 2025
    What to do if channel or needle is not recognized by BTG Laboratory Equipment?
    • H
      Heather SloanAug 26, 2025
      Check that the corresponding Channel Locking Bar is fully in the locked position. Ensure at least one needle is inserted into a channel for the channel to be available. If using a needle with a memory chip, try a new needle. If the channel may be defective, do not use it; move the needle(s) to a different channel and conduct the Needle Integrity and Functionality Test again.
  • J
    Jimmy HallAug 25, 2025
    What to do if BTG Laboratory Equipment doesn't recognize USB flash drive?
    • A
      Abigail BellAug 25, 2025
      Ensure the USB flash drive is connected to the port with the USB port icon. Try removing the USB flash drive, waiting a few seconds, and reconnecting it. If the problem persists, try using a different Galil Medical supplied USB flash drive or replace the USB flash drive with a new one.

Summary

1 SYSTEM OVERVIEW

1.2 Intended Use

Outlines the medical fields and tissue types the system is designed for.

1.6 Warnings

Provides critical safety warnings and precautions for operating the system.

2 SYSTEM DESCRIPTION

2.1 Touch Screen Monitor

Describes the primary interface for controlling and monitoring procedures.

2.9 Needle Connection Panel

Details the panel where cryoablation needles are connected.

3 NAVIGATING THE USER INTERFACE

3.1 Login Screen

Describes the initial screen for user authentication and system access.

3.3 Procedure Screen

The main operational screen for controlling cryoablation procedures.

4 SYSTEM OPERATION

4.1 Preparation for Use

Outlines the steps for setting up the system before a procedure.

4.2 Performing a Cryoablation Procedure

Details the step-by-step process of conducting a cryoablation treatment.

4.4 System Shutdown

Provides instructions for safely powering down and preparing the system for storage.

5 ADMINISTRATIVE FUNCTIONS

5.1 Configure Settings

Allows customization of system parameters, user accounts, and units.

6 SYSTEM CARE and MAINTENANCE

6.2 Cleaning

Details procedures for cleaning the system components and stylus.

6.3 Service and Preventive Maintenance

Outlines requirements for system servicing and preventive maintenance.

7 TROUBLESHOOTING

7.1 Software Recovery

Explains how to restore the system software to a previous version.

7.2 Electronics, Electrical and User Error Related Issues

Addresses common problems with system components and user errors.

7.3 Gas Issues

Covers troubleshooting steps for problems related to gas supply and connections.

8 SYSTEM SPECIFICATIONS

8.1 Operating Conditions

Specifies the recommended temperature and humidity for system operation.

8.4 Mechanical Specifications

Lists physical dimensions, weight, and capacity limits of the system.

9 REFERENCES

9.3 Contact

Provides contact information for Galil Medical Inc.

10 DISCLAIMER OF WARRANTY