INTELLIDOX DOCKING MODULE USER MANUAL || PREPARE MODULES FOR USE
BW TECHNOLOGIES BY HONEYWELL PAGE 36 OF 103
Connecting a Calibration Gas Cylinder
1. Connect a demand flow regulator to the calibration gas cylinder, and then attached
the supplied connector with attached 3/16 inch (9.5 millimeter) I.D. tubing to the
demand flow regulator.
2. Cut 1/8 inch (3.2 millimeter) I.D. tubing to an appropriate length. BW Technologies
recommends that the tubing is between 39 inches (1 meter) and 33 feet (10 meters)
in length.
3. Attach one end of the 1/8 inch (3.2 millimeter) I.D. tubing to the demand flow
regulator.
4. Insert a quick connect fitting into the free end of the tubing, and then connect it to Inlet
1 on the module.
5. If a multi-inlet key is inserted in an individual module or one or more connected
modules in a gang, then repeat steps 1, 2, 3 and 4 as necessary until up to 4
cylinders are attached.
6. Configure the gas inlets before using the module. See Configure Gas Inlets on
page 41.