INTELLIDOX DOCKING MODULE USER MANUAL || TROUBLESHOOTING
BW TECHNOLOGIES BY HONEYWELL PAGE 94 OF 103
Detectors
The detector firmware is not updated after a new firmware update file is
transferred via USB flash drive.
The file name is incorrect.
The firmware update file name is not the same as the downloaded file name.
Rename the file name to match the name of the downloaded firmware update
file, or download a new firmware update file. See Update Firmware.
The file is not in the System folder on the USB flash drive.
Create a folder called System on the USB flash drive. Move the file to
the System folder. See Updating Module Firmware via USB Flash Drive.
Update Detector Firmware is not enabled via Fleet Manager II software.
Use software to enable detector firmware updates. Transfer the modified
configuration file to the appropriate modules.
The firmware on the detector is the same version as the firmware update.
Verify that the firmware file on the USB flash drive is the most recent firmware
version. See Update Firmware.
The detector time is incorrect, or is not the same as the module time.
Use software to enable the Synchronize detector time and date option.
See Adjust Time and Date Settings via Fleet Manager II Software and in
the Automated Operations section, select Synchronize detector time and date.
The module time was not set via Fleet Manager II.
Use software to set the module time. See Adjust Time and Date Settings via
Fleet Manager II Software.
The time zone is incorrect.
Use software to select the time zone. See Adjust Time and Date Settings via
Fleet Manager II Software.
Daylight Savings Time did not work
The module Daylight Savings Time was not set.
Use software to automatically adjust clock for Daylight Savings Time. See Adjust
Time and Date Settings via Fleet Manager II Software.
The time zone is incorrect.
Use software to select the time zone. See Adjust Time and Date Settings via
Fleet Manager II Software.
The module time was not set via Fleet Manager II.
Use software to set the module time. See Adjust Time and Date Settings via
Fleet Manager II Software.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BW Technologies IntelliDoX and is the answer not in the manual?

BW Technologies IntelliDoX Specifications

General IconGeneral
BrandBW Technologies
ModelIntelliDoX
CategoryControl Unit
LanguageEnglish