INTELLIDOX DOCKING MODULE USER MANUAL || MAINTENANCE
BW TECHNOLOGIES BY HONEYWELL PAGE 86 OF 103
Calibration Equipment and Gases
The use of calibration gas cylinders other than those specified by BW Technologies
may result in unsafe calibration or possible non-recoverable failure of the equipment,
and will invalidate the warranty.
A demand flow regulator must be used with all calibration gas cylinder connections.
The module inlet should not be pressurized.
To receive a complete list of calibration gases and accessories approved for use with this
product, contact BW Technologies or an authorized distributor, or visit the Order
Information section of the product website at www.honeywellanalytics.com.
Sticky gases (CL
2
, NH
3
and NO
2
) calibration and bump tests of are only allowed on first
bay. The user cannot perform calibration and bump tests from bay-2 to bay-5. And for that
configured inlet, IDOX will display as .