8545-M838-O1 Rev D 825 Installation & Technical
SETUP AND CALIBRATION, CONT.
Total - Calibration – Options
Sample Rate:
With the Total – Calibration – Options screen displayed the current setting for the
Sample Rate: parameter will be shown. If the setting displayed is acceptable, proceed
to the next parameter.
Otherwise, use the numeric keys to enter the sample rate in samples per seconds and
then press the Down Arrow to advance to the next parameter or press the ENTER key
to save the new setting and proceed to the next screen. Allowable values are 1 to 200.
NOTE: The Sample Rate: will automatically be set to a 2 when Minimal Filtering:
has been selected, or a 1 when Moderate has been selected for Filtering:.
Num Cal Pts:
Calibration of the scale can be performed using multiple calibration points to obtain a
more precise weight reading over the entire span of the scale.
With the Total – Calibration – Options screen displayed the current setting for the
Num Cal Pts parameter will be shown. If the setting displayed is acceptable, proceed
to the next parameter.
Otherwise, use the numeric keys to enter the number of calibration points and then press
the Down Arrow to advance to the next parameter or press the ENTER key to save the
new setting and proceed to the next screen. Allowable values are 1 to 4.
NOTE: The Num Cal Pts parameter is used to set the calibration points in
addition to the zero (no load) weight calibration point. DO NOT include the
zero (no load) weight calibration point when selecting the number of calibration
points. For example, selecting 2 for the Num Cal Pts parameter, you would
have a total of three calibration points, the zero (no load) weight and two
calibrations points that could correspond to a mid-point weight and the test
(load) weight.