Chapter 5 – Troubleshooting
Section 5.3 – Error Messages
5 - 10
Bulbs Message This occurs if: Service Light
The bulbs are not
working properly in
chamber(s) (1, 2 or
both). Treatment has
been stopped in
chamber(s) (1, 2 or
both).
Bulbs need to be replaced.
Contact your authorized service
representative for assistance.
On after
‘Continue’ is
selected
The illumination sensors
are not working properly.
Treatment must be
canceled.
There are issues with the
illumination sensors. Contact
your authorized service
representative for assistance.
On after
‘Cancel’ is
selected
There is an issue with the
illumination sensors.
Treatment must be
canceled. Contact
service for assistance.
There are issues with the
illumination sensors. Contact
your authorized service
representative for assistance.
On after
‘Cancel’ is
selected
A Treatment Issue has
occurred (T1004).
Replace bulbs ‘X’ and
‘Y.’ Refer to the
Operator’s Manual for
assistance.
The listed bulbs need to be
replaced. Contact your
authorized service
representative for assistance.
Off
An Integrity Check Issue
has occurred (C0011).
Contact service for
assistance.
There are issues with the
illumination sensors. Contact
your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred. (C0012).
Contact service for
assistance.
There are issues with the
illumination sensors. Contact
your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0013).
Replace bulbs ‘X’ and
‘Y.’ Refer to the
Operator’s Manual for
assistance.
The listed bulbs need to be
replaced. Contact your
authorized service
representative for assistance.
On