Page #38 background imageLoading...
Page #38 background image
36 3 Interface
3
M Format
#zz_CC_RR_J_±LLL.LLL_±PPP.PPP_±DDD.DDD_±KKK.KKK_HHH.H_III.I_www.w_NNN.N_OOO.O_QQQ.Q_vvv.v_S
SS.S_TTT.T_UUU.U_VVV.V_WWW.W_xxx.x_ll_pp_kk
_±ccc.ccc_±ddd.ddd_±hhh.hhh_±aaa.aaa_±bbb.bbb_±iii.iii_±EE
E.EEE_±FFF.FFF_e_ff_gg_jj_±mmm.mmm_±nnn.nnn_±ooo.ooo_±qqq.qqq_±rrr.rrr_±sss.sss_tt_uu_±YYY.YYY_±ZZZ.
ZZZ_YYMMDD_HHMMSS:GG CR
Data field
Data type
Unit
Min.
Max.
Note
Tester ID
z
2-digit decimal
--
00
99
CH#
C
2-digit decimal
--
00
32
Error code
R
2-digit decimal
--
01
18
*1
Result J
One ASCII code
(1-digit hexadecimal)
-- 1 D
1: LL FAIL
2: Pass
4: UL FAIL
9: LL2 FAIL
C: UL2 FAIL
D: Error
Leak
L
Fixed point
Leak unit
-999.000
+999.000
Test Pressure
P
Fixed point
Test press unit
-999.000
+999.000
ΔP
D
Fixed point
Pa
-999.999
+999.999
K(Ve)
K
Fixed point
K(Ve) unit
-999.999
+999.999
DL1
H
Fixed point
Second
000.0
999.9
CHG
I
Fixed point
Second
000.0
999.9
BAL1
w
Fixed point
Second
000.0
999.9
BAL2
N
Fixed point
Second
000.0
999.9
DET
O
Fixed point
Second
000.0
999.9
BLW
Q
Fixed point
Second
000.0
999.9
END
v
Fixed point
Second
000.0
999.9
EXH
S
Fixed point
Second
000.0
999.9
MB1
T
Fixed point
Second
000.0
999.9
MB2
U
Fixed point
Second
000.0
999.9
PCHK
V
Fixed point
Second
000.0
000.2
*2
PCHG
W
Fixed point
Second
000.0
999.9
PEXH
x
Fixed point
Second
000.0
999.9
Leak unit
l
2-digit decimal
--
00
16
*3
Test Press Unit
p
2-digit decimal
--
00
08
*3
K(Ve) Unit
k
2-digit decimal
--
00
03
*3
BAL (UL)
c
Fixed point
Leak unit
-999.999
+999.999
BAL (LL)
d
Fixed point
Leak unit
-999.999
+999.999
DET (UL2)
h
Fixed point
Leak unit
-999.999
+999.999
DET (UL)
a
Fixed point
Leak unit
-999.999
+999.999
DET (LL)
b
Fixed point
Leak unit
-999.999
+999.999
DET (LL2)
i
Fixed point
Leak unit
-999.999
+999.999
TP UL
E
Fixed point
Test press unit
-999.999
+999.999
TP LL
F
Fixed point
Test press unit
-999.999
+999.999
Press Monitor
e
1-digit decimal
--
0
1
Comp Type
f
2-digit decimal
--
00
02
*3
Mastering Iterations
g
2-digit decimal
--
01
20
Number of Samples
j
2-digit decimal
--
00
20
Drift Comp.
m
Fixed point
Leak unit
- 999.999
+ 999.999
Mastering Comp
n
Fixed point
Leak unit
- 999.999
+ 999.999
D. Comp Upper Limit
o
Fixed point
Leak unit
- 999.999
+ 999.999
D. Comp Lower Limit
q
Fixed point
Leak unit
- 999.999
+ 999.999
M. Comp Upper Limit
r
Fixed point
Leak unit
- 999.999
+ 999.999
M. Comp Lower Limit
s
Fixed point
Leak unit
- 999.999
+ 999.999
No item
t
2-digit decimal
--
00
00
Fixed to 00
NR Iterations
u
2-digit decimal
--
01
20
EP Precharge
Y
Fixed point
Test press unit
- 999.999
+ 999.999