Pre-use Checkout
Instructions for Use Apollo SW 4.5n 101
Part Number: 9053586, 3rd edition
Figure 45. Non-Critical Error During Self Test
Interruption of the test is symbolized by an
exclamation mark.
Red Operation of the anesthesia machine is
impossible or not permitted. The error must be
remedied and the test must be repeated.
The self-test can no longer be canceled at this
point.
WARNING!
Risk of device failure or patient injury
Functions coded yellow do not meet with the
specified technical data.
The error should be remedied as soon as
possible.
WARNING!
Risk of device failure or patient injury
Functions coded red must be remedied before
starting, e.g. if there is no O
2 supply.
The device cannot be operated in this state.
WARNING!
Risk of inadequate monitoring
If the flow sensor, oxygen sensor, or gas sensor
is not operational, adequate substitute
monitoring must be ensured before starting the
anesthesia machine!
Special attention is required if operation is
initiated.