System Setup
Instructions for Use Apollo SW 4.5n 85
Part Number: 9053586, 3rd edition
Fuses for auxiliary outlets
If a circuit breaker is tripped (position 0):
1. Remedy the fault.
2. Press the switch on the circuit breaker into
position 1.
The circuit breaker is active again.
In cases of a blown safety fuse:
1. Remedy the fault.
2. Have the safety fuse replaced by an expert.
Establishing potential equalization
e.g., for intracardiac or intracranial surgery.
Differences in electrical potential between devices
can be reduced by potential equalization.
Potential equalization does not replace the protective
ground connection.
During operation, the potential equalization
connectors must be readily accessible and the
connection must be able to be disconnected without
the use of tools.
WARNING!
Risk of electric shock
Connecting devices to the auxiliary outlets of
the anesthesia machine can cause an increase
in leakage current beyond permissible values if
the protective conductor of a device fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting a
device or several devices increases the leakage
current to a value which exceeds the
permissible value, do not use the auxiliary
outlets of the anesthesia machine. Instead, use
a mains power socket on the wall.
The system must fulfill the requirements for
medical electrical equipment in accordance with
the relevant standards, see “Relevant
standards” on page 308.