Pre-use Checkout
Instructions for Use Apollo SW 4.5n 91
Part Number: 9053586, 3rd edition
Overview
The pre-use checkout procedure must be performed
to ensure that the Apollo is ready for use. This is a
recommended procedure. Follow the institution’s
policies for specific procedures.
If the Apollo fails any checkout routine, do not use
the machine until corrective action is taken. If
indicated, contact an authorized representative of
DrägerService for inspection of the unit.
Checking the Workstation accord-
ing to the Check List
The pre-use checkout procedure consists of a
manual procedure performed by the user, followed
by an automated self test. The manual procedure is
summarized in the check list that is displayed after
the machine is powered on.
Prerequisites
The device has been prepared (see “Cleaning and
Maintenance” on page 225) and assembled ready for
operation.
The pipeline supply and the power supply must be
connected.
WARNING!
Risk of device failure and/or patient injury
Do not insert any additional components into or
modify the Apollo after the checkout procedure
has been started.
The anesthesia machine will not meet the
specified technical data.
WARNING!
Risk of patient injury
Inappropriate hose length affects compliance
and can result in incorrect tidal volume delivery
to the patient.
Patient hoses must be adjusted to the
appropriate lengths prior to performing the leak
and compliance tests.