Specifications
Instructions for Use Apollo SW 4.5n 295
Part Number: 9053586, 3rd edition
Frequency measurement
MVLEAK
Range 0to9.99L/min
Resolution of the indication 0.01 L/min
Accuracy ±15% of (MV
EXP + MVLEAK) or ±0.1 L/min, whichever is greater
MV
MAND
Range 0 to 99.9 L/min
Resolution of the indication 0.1 L/min
Accuracy ±8% of measured value or ±0.1 L/min, whichever is greater
MV
SPON
Range 0 to 99.9 L/min
Resolution of the indication 0.1 L/min
Accuracy ±8% of measured value or ±0.1 L/min, whichever is greater
O
2 Uptake
Range 0 to 9999 mL/min
Resolution of the indication 1 mL/min
Accuracy ±15% or ±20 mL/min, whichever is greater
MV
SPON
Range 0 to 9999 mL/min
Resolution of the indication 1 mL/min
Accuracy ±20% or ±20 mL/min, whichever is greater
Frequency (Freq.)
Range 1 to 100 bpm
Resolution of the display 1 bpm
Accuracy ±10% or ±1 bpm, whichever is lower (6 to 100 bpm);
±0.3 bpm (<6 bpm)

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
TypeAnesthesia Workstation
Gas SupplyO2, N2O, Air
Integrated MonitoringYes
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
VaporizerDräger Vapor 2000
Patient MonitoringECG, SpO2, NIBP, Temperature
Power Supply100 – 240 VAC, 50/60 Hz
Tidal Volume Range20-1500 ml
Vaporizer MountsUp to 2

Related product manuals