Specifications
Instructions for Use Apollo SW 4.5n 291
Part Number: 9053586, 3rd edition
Ventilator
(electronically controlled, electrically driven piston ventilator, fresh gas decoupled)
Ventilation modes Volume Mode, Pressure Mode, Volume AF (optional), CPAP (optional),
and Pressure Support mode (optional), synchronized volume and
pressure modes
Settings:
Pressure limitation P
MAX in Volume and
Volume AF Mode
(PEEP+10) to 70 cmH2O
Accuracy ±10% of set value or ±3 cmH
2O, whichever is greater
Inspiration pressure P
INSP in Pressure Mode (PEEP+5) to 70 cmH2O
Accuracy ±10% of set value or ±3 cmH
2O, whichever is greater
Tidal volume V
T (system compliance
compensated)
All values apply under ATPD conditions:
in Volume and Volume AF Mode 20 mL to 1400 mL
1)
with optional Pressure Support 5 mL to 1400 mL
1)
Accuracy (5 mL to 150 mL) ±10% of set value or ±10 mL, whichever is greater
Accuracy (over 150 mL) ±5% of set value or ±15 mL, whichever is greater
Frequency 3 to 100 bpm
Accuracy ±10% of set value or ±1 bpm, whichever is lower
Frequency
MIN
in Pressure Support Mode 3 to 20 bpm or “OFF”
in Pressure Support CPAP Mode “OFF”
Accuracy ±10% of set value or ±1 bpm, whichever is lower
T
INSP 0.2 seconds to 6.7 seconds
Insp./exp. time ratio I:E Derived from frequency and TINSP
Range: 5:1 to 1:99
Inspiration pause T
IP : TINSP 0 % to 60%
1) Due to gas measurement sampling, leakage (both at the patient and in the device), and resistance/compliance of the patient and the
breathing circuit, the maximum delivered tidal volume may be limited.