288 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Auxiliary O2 flow meter
All values apply under STPD conditions:
Breathing system
Fresh-gas flow, total flow meter 0 to 10 L/min
Accuracy ±10% of the max. displayed value for 50% O
2 and 50% N2O
calibrated at 5 L/min
Resolution 0.5 L/min (from 0.5 to 2 L/min)
1.0L/min (from 2to10L/min)
O
2 flush >35 L/min
Standardized to ATPD conditions.
Connection Staged connector for use with various hose diameters
Fresh-gas flow 0 to 10 L/min
Accuracy of the flow display ± 10% of full scale
Resolution of the display 0.5 L/min
The auxiliary O
2 flow meter is not pressure monitored; this monitoring
must be ensured by the connected device.
Total gas volume Without breathing hoses, incl. absorber
in Man/Spont typically 3.7 L
in automatic mode typically 4.0 L (incl. piston volume)
Compliance
All values apply under STPD conditions:
Without breathing hoses, flexible bag arm
in Man/Spont typically 3.7 mL/cmH
2O
in automatic mode typically 2.3 mL/cmH
2O
Absorber volume
Reusable absorber canister, filled 1.5 L
CLIC absorber (Drägersorb 800 +) 1.2 L
CLIC absorber (Drägersorb Free) 1.2 L