EasyManuals Logo

Dräger Apollo User Manual

Dräger Apollo
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #288 background imageLoading...
Page #288 background image
Specifications
288 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Auxiliary O2 flow meter
All values apply under STPD conditions:
Breathing system
Fresh-gas flow, total flow meter 0 to 10 L/min
Accuracy ±10% of the max. displayed value for 50% O
2 and 50% N2O
calibrated at 5 L/min
Resolution 0.5 L/min (from 0.5 to 2 L/min)
1.0L/min (from 2to10L/min)
O
2 flush >35 L/min
Standardized to ATPD conditions.
Connection Staged connector for use with various hose diameters
Fresh-gas flow 0 to 10 L/min
Accuracy of the flow display ± 10% of full scale
Resolution of the display 0.5 L/min
The auxiliary O
2 flow meter is not pressure monitored; this monitoring
must be ensured by the connected device.
Total gas volume Without breathing hoses, incl. absorber
in Man/Spont typically 3.7 L
in automatic mode typically 4.0 L (incl. piston volume)
Compliance
All values apply under STPD conditions:
Without breathing hoses, flexible bag arm
in Man/Spont typically 3.7 mL/cmH
2O
in automatic mode typically 2.3 mL/cmH
2O
Absorber volume
Reusable absorber canister, filled 1.5 L
CLIC absorber (Drägersorb 800 +) 1.2 L
CLIC absorber (Drägersorb Free) 1.2 L

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals