Configuration
212 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
If set to No, MV alarms and flow apnea monitoring
are deactivated in bypass mode.
Device related
See 2 in Figure 131.
CAUTION!
Risk of insufficient monitoring
If the setting MV alarms in Bypass mode? is set to
no, MV alarms and flow apnea monitoring are
deactivated.
Special attention is required.
Device-related alarm limits Factory setting
O
2 cylinder low alarm active
at: – –
1)
, 150 to 450
Determine the pressure at
which the warning
O2 CYLIND. LOW is to be
issued. This menu item only
appears if the O
2 cylinder has
been configured as gas supply,
see page 215
1) The function is disabled.
450 psi
CO
2 absorber time limit:
1 to 60 days; – –
1)
Determine the operating time
interval for the soda lime. As
soon as the interval has
expired the advisory message
CO
2 ABSORB. DEPLETED?
will be issued.
7 days