Introduction
12 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Not for use in areas of explosion hazard
Safe connection with other electrical equipment
For further information, see “Device combinations” on
page 306.
WARNING!
Risk of explosion and fire
The medical device is not approved for use in
areas where oxygen concentrations greater than
25 Vol%, or combustible or explosive gas
mixtures are likely to occur.
WARNING!
Risk of electric shock or device malfunction
Any connected devices or device combinations
not complying with the requirements mentioned
in these instructions for use can compromise the
correct functioning of the medical device and lead
to an electric shock.
Electrical connections to equipment not listed
in these instructions for use or these assembly
instructions must only be made when approved
by each respective manufacturer.
Before operating the medical device, strictly
comply with the instructions for use of all
connected devices or device combinations.
WARNING!
Risk of electric shock
A test for leakage current must be performed by
qualified biomedical engineering personnel
before use if the Apollo is interfaced with other
equipment.
WARNING!
Risk of explosion, fire
If an oxygen leak is suspected within or near the
anesthesia machine, do not initiate operation.
Disconnect all oxygen supplies and contact a
trained service technician.
WARNING!
Risk of use error
Various potentially dangerous situations may
occur which demand the attention of trained
personnel.
The workstation may only be used under the
permanent supervision of qualified medical
personnel so that assistance can be provided
immediately in the event of any malfunctions.
WARNING!
Risk of fire
The flow sensor can ignite medications or other
substances based on highly flammable
substances.
Do not nebulize medications or other
substances that are easily flammable or spray
them into the device.
Do not use substances containing alcohol.
Do not allow flammable or explosive
substances to enter the breathing system or the
breathing circuit.
Do not use cyclopropane or ether.
WARNING!
Risk due to failure of flow measurement
Deposits that were not removed during
reprocessing can damage the measuring wires in
the flow sensor or cause a fire.
Before inserting the flow sensor check for
visible damage, soiling, and particles. Repeat
this check regularly.
Replace flow sensors when damaged, soiled, or
not particle-free.
WARNING!
Risk of device failure and/or danger to patient
Magnetic fields may negatively influence the
proper function of the medical device, thus
endangering the patient or user.
The medical device must not be used in the
vicinity of magnetic resonance imagers (MRI,
NMR, NMI).

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals