Troubleshooting
Instructions for Use Apollo SW 4.5n 263
Part Number: 9053586, 3rd edition
Fresh-gas delivery failure
Figure 178. Gas Mixer Failure If the fresh-gas mixer fails, the following message is
displayed in the status field on the screen
(1 in Figure 178):
GAS MIXER FAIL
In addition the numerical values for the fresh-gas
flows appear grayed out to indicate that they may be
inaccurate.
The current ventilation mode remains active.
1. Close N
2O and Air flow control valves.
2. Check vaporizer setting.
3. Set O
2 flow to the desired level using the total
flow meter.
Note: Refer to the total flow meter for approximate
flow (see the chapter “Specifications” of the
instructions for use for accuracy).
1
WARNING!
Risk of patient injury
If the fresh-gas delivery fails, the anesthesia
machine automatically discontinues the fresh-
gas flow. An O
2 flow must be delivered to the
patient.
Check vaporizer setting and set the O
2 flow to a
sufficient level.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
TypeAnesthesia Workstation
Gas SupplyO2, N2O, Air
Integrated MonitoringYes
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
VaporizerDräger Vapor 2000
Patient MonitoringECG, SpO2, NIBP, Temperature
Power Supply100 – 240 VAC, 50/60 Hz
Tidal Volume Range20-1500 ml
Vaporizer MountsUp to 2

Related product manuals