Specifications
292 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Measuring systems
Inspiration flow Derived from VT and TINSP
in Volume Mode 0.1 to 100 L/min ±10%
in Volume AF Mode max. 150 L/min +10%
in Pressure Mode max. 150 L/min +10%
PEEP
in Volume and Volume Mode 0 to 20 cmH
2O (max. PMAX - 10 cmH2O)
in Pressure and Pressure Support Mode 0 to 20 cmH
2O (max. PINSP - 5 cmH2O)
Accuracy ±10% of set value or ±2 cmH
2O, whichever is greater
1)
ΔPPS
in Volume, Volume AF and Pressure
Modes
3to50cmH2O (max. PINSP)
in Pressure Support 3 to 50 cmH2O (max. PINSP); 0to2cmH2O = Pressure Support CPAP
Trigger 0.3to15L/min orOFF
T
SLOPE
in Pressure Mode, Volume AF Mode and
Pressure Support Mode
0 seconds to 2 seconds
1) Due to gas measurement sampling and leaks (both at the patient and in the device), the end-expiratory PEEP value may be lower than specified at the end of
long expiratory phases.
Pressure Measurement
(piezo-resistive)
Respiratory pressure
Range 20to99cmH
2O
Resolution of the measurement 0.1 cmH2O
Accuracy ±4% of the measured value or ±2 cmH
2O, whichever is greater

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals