Monitoring
182 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
3. Change the application site after not more than 4
hours in order to avoid a build-up of blood
pressure (blocked circulation).
Follow the specific Instructions for Use when using
other Nellcor sensors!
Test considerations and oximeter accu-
racy
Functional testers and patient simulators
For functional testing of the pulse oximeter sensors
and cables with a functional tester or patient
simulator follow the individual testing device's
operator's manual, especially regarding the suitability
and accuracy of the simulated values.
CAUTION!
Risk of inaccurate data
If simulators are used as calibrators, the SpO
2
module may produce incorrect data.
Simulators must not be used as calibrators.