EasyManuals Logo
Home>Dräger>Medical Equipment>Apollo

Dräger Apollo User Manual

Dräger Apollo
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #260 background imageLoading...
Page #260 background image
Troubleshooting
260 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
4. Ventilate the patient manually.
5. Ensure adequate substitute monitoring.
Note: If a D-Vapor is in use and a power failure
occurs, refer to the Instructions for Use of the
D-Vapor for a description of system behavior
in a power fail situation.
Note: Refer to the total flow meter for approximate
flow (see the chapter “Specifications” of the
instructions for use for accuracy).
Gas failure
Figure 175. NO O2 SUPPLY Alarm MessageIf the gas supply fails, the Apollo displays a
corresponding message in the status field at the top
of the screen (1 in Figure 175):
NO AIR SUPPLY, NO N
2O SUPPLY, or
NO O
2 SUPPLY
1. Open the valve on the corresponding backup
cylinder at the back of the machine
(see Figure 176).
2. Restore the pipeline supply.
If there is no cylinder backup supply for the failed
gas, the corresponding LED on the front panel of the
machine will flash red.
In case of N
2O or air failure, 100% O2 should be
delivered; be sure to set the O
2 to an appropriate
flow.
WARNING!
Risk of patient injury
If all power supplies fail, the screen display will
be dark and automatic ventilation will cease.
The patient must be ventilated manually.
1
CAUTION!
Risk of patient injury
In case of a gas supply failure, adjust the gas
concentration manually.
Note the contra-indications for 100 % O
2.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals