Cleaning and Maintenance
Instructions for Use Apollo SW 4.5n 255
Part Number: 9053586, 3rd edition
Replacing the water trap
The water trap must be replaced under any of the
following conditions:
It becomes severely soiled.
–The WATER TRAP SAMPL. LINE? alarm
message persists even after the water trap has
been drained (with the sample line correctly
installed and free of any blockage).
The water trap has been in use for its maximum
life of four weeks.
1. Pull the old water trap off to the front, dispose of
with domestic waste as shown in Figure 172.
Observe the hospital's hygiene regulations.
2. Mark the new water trap with the current date.
Use the space provided for this purpose.
3. Push the new water trap into place until it
engages.
Disposing of the medical device
At the end of its service life:
Have the medical device appropriately disposed
of in accordance with applicable laws and
regulations.
WARNING!
Risk of gas measurement failure and device
failure
If the water trap is used longer than intended,
the diaphragm may become brittle and allow
water and bacteria to enter the measurement
system. Such contamination affects the gas
measurement which may fail as a result.
The water trap must be replaced at least every
four weeks.
WARNING!
Risk of infection
The device and its components must be
disinfected and cleaned before disposal!

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
TypeAnesthesia Workstation
Gas SupplyO2, N2O, Air
Integrated MonitoringYes
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
VaporizerDräger Vapor 2000
Patient MonitoringECG, SpO2, NIBP, Temperature
Power Supply100 – 240 VAC, 50/60 Hz
Tidal Volume Range20-1500 ml
Vaporizer MountsUp to 2

Related product manuals