Specifications
286 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Protection class
Maximum oxygen inlet flow
(including auxiliary O2 flow meter and O2
flush)
72 L/min
Scavenging flow for anesthetic gas
receiving system
30 to 50 L per minute
Dew point >41 °F (>5 °C) below ambient temperature
Oil content <0.1 mg/m
3
Particles Dust-free air (filtered with pores <1 µm)
Driving gas consumption None
Noise emissions from device Free field measurements complying with ISO 3744
Average sound pressure level Leq(A)
during ventilation with typical settings
42 dB(A)
Sound pressure level L(A) of the alarm
tones at the workstation, measured
according to IEC 60601-1-8 Edition 2.1:
The alarm volume is adjustable, refer to “Alarm volume” on page 221.
Alarm volume (high priority) Approximately 48 to 70 dB(A)
Alarm volume (medium priority) Approximately 48 to 70 dB(A)
Alarm volume (low priority) Approximately 48 to 70 dB(A)
Auxiliary alarm
55 dB(A) and 65 dB(A)
Mains supply failure alarm 55 dB(A) and 65 dB(A)
Noise emission in normal operation with
ventilation (breathing sound set to OFF)
Leq (5 cycles)
45 dB(A)
Anesthesia machine I, in accordance with IEC 60601-1
IP class as per IEC 60529 IPX0
SpO
2 sensor Type BF
electrically isolated from protective conductor