Pre-use Checkout
106 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Additional suggestions to isolate compo-
nents of the breathing system for leaks:
Carry out the described measures:
Patient Sample Line Isolation:
1. Remove the sample line for gas measurement
and seal the Luer Lock connection on the
Y-piece.
2. Perform leak test.
Exclude the breathing hoses from the leak test
1. Remove the patient circuit from the breathing
system.
2. Install a leak-free hose between the inspiration
and expiration ports. The breathing bag must be
on the bag arm.
3. Perform leak test.
Isolation of Vaporizers:
1. Remove the vaporizers from the anesthesia
machine.
2. Perform leak test.
Note: The self test does not check for internal
vaporizer leakage; after filling or changing
vaporizers, perform the Standby leak test on
each vaporizer (see page 118).

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
TypeAnesthesia Workstation
Gas SupplyO2, N2O, Air
Integrated MonitoringYes
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
VaporizerDräger Vapor 2000
Patient MonitoringECG, SpO2, NIBP, Temperature
Power Supply100 – 240 VAC, 50/60 Hz
Tidal Volume Range20-1500 ml
Vaporizer MountsUp to 2

Related product manuals