Specifications
Instructions for Use Apollo SW 4.5n 293
Part Number: 9053586, 3rd edition
PPEEP, PPEAK, PPLAT, PMEAN
Range 20to99cmH2O
Resolution of the display 1 cmH2O
Accuracy ±4% of the measured value or ±2 cmH
2O, whichever is greater
Respiratory pressure at auxiliary common gas
outlet (Aux CGO)
Range 20to99cmH
2O
Resolution of the measurement 0.1 cmH
2O
Accuracy ±8% of the measured value or ±3 cmH2O, whichever is greater
P
PEAK, PMEAN at auxiliary common gas outlet
(Aux CGO)
Range 20to99cmH2O
Resolution of the display 1 cmH
2O
Accuracy ±8% of the measured value or ±3 cmH
2O, whichever is greater
Central supply pressure
Range 0 psi to 140 psi
Resolution of the display 1.5 psi
Accuracy ±4% or ±3 psi
Cylinder pressure
(applies for Silverline pressure regulators)
Range 0 psi to 3600 psi
Resolution of the display 14 psi
Accuracy ±4% or ±87 psi
Pressure Measurement
(pressure indicator, dial type)
Range 20to80cmH
2O
Resolution of the display 5 cmH2O
Accuracy ±5% of the measured value or ±2 cmH
2O, whichever is greater