Specifications
294 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Flow measurement
(hot wire anemometry)
All values apply under ATPD conditions:
Flow
Range –180 to 180 L/min
Resolution of the measurement 0.1 L/min
Accuracy at 60 L/min ±8% of measured value
Tidal volume V
T
Range 0 mL to 9999 mL
Resolution of the display 1 mL
Accuracy ±8% of the measured value or ±5 mL, whichever is greater
Delta V
T
Range 0 mL to 9999 mL
Resolution of the display 1 mL
Accuracy ±16% or ±10 mL, whichever is greater
Volume VT
INSP
Range 0 mL to 9999 mL
Resolution of the display 1 mL
Accuracy ±8% of the measured value or ±5 mL, whichever is greater
Minute volume MV
Range 0 to 99.9 L/min
Resolution of the display 0.1 L/min
Accuracy ±8% of the measured value or ±0.1 L/min, whichever is greater
Compliance C
PAT
Range 0 to 250 mL/cmH2O
Resolution of the display 0.1 mL/cmH
2O
Accuracy ±15% of the measured value or ±0.5 mL/cmH2O, whichever is greater