Specifications
Instructions for Use Apollo SW 4.5n 289
Part Number: 9053586, 3rd edition
Flexible arm for breathing bag
Volume 0.13 L
Compliance 0.13 mL/cmH
2O
Total system leakage (as per ISO 8835-2) <150 mL/min at 30 cmH
2O
Standardized to BTPS conditions.
Pressure limitation valve APL
Adjustment range 5 to 70 cmH
2O
Accuracy between 5 and 15 L/min ±15% of set value or ±3 cmH
2O, whichever is greater
Pressure drop at 30 L/min 2.8 cmH2O (in Spont position; wet and dry)
Resistance Reusable absorber or CLIC absorber, normal operation (filled with
Drägersorb 800 +)
With standard bag tube With flexible bag arm
Inspiratory Expiratory Inspiratory Expiratory
As per ISO 8835-2, dry, max. ±6 cmH
2O,
with hose set for adults M30146
-4.4 cmH2O 4.2 cmH2O-4.6cmH2O 4.2 cmH2O
As per ISO 8835-2, dry, sole breathing
system without patient hoses
-3.0 cmH
2O 3.0 cmH2O-3.3cmH2O 3.3 cmH2O
Minimal Limited Pressure (in accordance with
ISO 8835-5 and ISO 80601-2-13)
-3 cmH
2O