EasyManua.ls Logo

Dräger Apollo

Dräger Apollo
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Instructions for Use Apollo SW 4.5n 289
Part Number: 9053586, 3rd edition
Flexible arm for breathing bag
Volume 0.13 L
Compliance 0.13 mL/cmH
2O
Total system leakage (as per ISO 8835-2) <150 mL/min at 30 cmH
2O
Standardized to BTPS conditions.
Pressure limitation valve APL
Adjustment range 5 to 70 cmH
2O
Accuracy between 5 and 15 L/min ±15% of set value or ±3 cmH
2O, whichever is greater
Pressure drop at 30 L/min 2.8 cmH2O (in Spont position; wet and dry)
Resistance Reusable absorber or CLIC absorber, normal operation (filled with
Drägersorb 800 +)
With standard bag tube With flexible bag arm
Inspiratory Expiratory Inspiratory Expiratory
As per ISO 8835-2, dry, max. ±6 cmH
2O,
with hose set for adults M30146
-4.4 cmH2O 4.2 cmH2O-4.6cmH2O 4.2 cmH2O
As per ISO 8835-2, dry, sole breathing
system without patient hoses
-3.0 cmH
2O 3.0 cmH2O-3.3cmH2O 3.3 cmH2O
Minimal Limited Pressure (in accordance with
ISO 8835-5 and ISO 80601-2-13)
-3 cmH
2O

Table of Contents