EasyManuals Logo

Dräger Apollo User Manual

Dräger Apollo
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
Pre-use Checkout
108 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Apollo is ready for operation about 1 minute after
initiating. The O
2 sensor is completely calibrated
after about 5 minutes.
After the minimal self test, the anesthesia machine
switches to Standby.
Note: The >Cancel Test< key is also available in
the self test screen.
WARNING!
Risk of device failure and/or patient injury
Canceling the self test may lead to
malfunctions; greater attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency situation, carry out a complete
self test as soon as practicable.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
TypeAnesthesia Workstation
Gas SupplyO2, N2O, Air
Integrated MonitoringYes
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
VaporizerDräger Vapor 2000
Patient MonitoringECG, SpO2, NIBP, Temperature
Power Supply100 – 240 VAC, 50/60 Hz
Tidal Volume Range20-1500 ml
Vaporizer MountsUp to 2

Related product manuals