Pre-use Checkout
108 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Apollo is ready for operation about 1 minute after
initiating. The O
2 sensor is completely calibrated
after about 5 minutes.
After the minimal self test, the anesthesia machine
switches to Standby.
Note: The >Cancel Test< key is also available in
the self test screen.
WARNING!
Risk of device failure and/or patient injury
Canceling the self test may lead to
malfunctions; greater attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency situation, carry out a complete
self test as soon as practicable.