Ventilation
144 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Note: Pressure Support mode can also be started
without presetting values:
– Press the >Press. Supp< key located at
the bottom of the display panel, and press
the rotary knob. Default parameter values
are used.
Note: There is 15 second timeout period for
making ventilation mode changes, with a
5-second audible tone sequence after the
first 10 seconds. If the new setting is not
confirmed within the timeout period, the
current ventilation setting remains in effect.