Configuration
216 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
–TINSP changes with Freq. if synchronization is
off: Yes or No
When >Yes< is set:
T
INSP is automatically adjusted when the
frequency is changed so that the ratio of
inspiration to expiration I:E remains constant.
This only applies if synchronization has not been
set.
When >No< is set:
T
INSP remains independent of the change in
frequency and the ratio of inspiration to expiration
I:E changes accordingly.
Factory setting: >Yes<
–P
AW low alarm limit adjusts with PEEP: Yes or
No
When >Yes< is set:
The low alarm limit for airway pressure (P
AW) will
be automatically changed when the PEEP value
is changed.
When >No< is set:
The low alarm limit for airway pressure (P
AW) will
be unaffected by changes in the PEEP value.
In the Pressure Mode, the P
AW low alarm limit will
not exceed the P
INSP - 2. This is also true for
changes to P
INSP.
In the Pressure Support Mode, PEEP + ΔP
PS -2
will not be exceeded. This is also true for
changes to ΔP
PS.
Factory setting: >No<

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals