Cleaning and Maintenance
Instructions for Use Apollo SW 4.5n 253
Part Number: 9053586, 3rd edition
Technical customers documentation according to
IEC/EN 60601 is available upon request.
Repair
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
Filter mat, power supply Every 12 months Replace Service personnel
Dust filter, ventilator unit Every 12 months Replace Service personnel
O-rings, vapor plug system Every 12 months Replace Service personnel
O-rings for holder, water trap Every 12 months Replace Service personnel
Nafion hose on patient gas module Every 12 months Replace Service personnel
Filter mat, housing cover Every 2 years Replace Service personnel
Sintered filter, gas inlet Every 2 years Replace Service personnel
PEEP diaphragm, breathing sys-
tem
Every 2 years Replace Service personnel
Man.Spont.-Automatic reversing
diaphragm
Every 2 years Replace Service personnel
Bacterial/viral filter, patient gas
module
Every 2 years Replace Service personnel
Bacterial/viral filter, ventilation and
gas controller
Every 2 years Replace Service personnel
O-rings between valve plate and
diaphragm cover of breathing sys-
tem
Every 2 years Replace Service personnel
Lower diaphragm of ventilator +
O-ring
Every 3 years Replace Service personnel
Lead gel battery in UPS Every 3 years Replace Service personnel
Or when message BATTERY
LOW is displayed.
Replace Experts
Silverline pressure regulator After 6 years Inspection and
service
Experts
Component Interval Measure Personnel responsible