Specifications
Instructions for Use Apollo SW 4.5n 297
Part Number: 9053586, 3rd edition
O2 Measurement (consumption-free, paramagnetic measurement)
Measuring range 0 to 100 Vol.%
Resolution of the measurement 0.1 Vol.%
Resolution of the indication
(for ins. O
2, exp. O2)
1Vol.%
Accuracy ±(2.5 Vol% +2.5 % rel.)
CO
2 Measurement (infrared spectrometry)
Measuring range 0 to 13.6 Vol%
0 to 13.6 kPa
0 to 102 mmHg
(at an ambient pressure of 1013 hPa / 760 mmHg)
Resolution of the measurement 1 mmHg
Resolution of the indication
(for inCO
2, etCO2)
1 mmHg
Accuracy ±(0.43 Vol% +8 % rel.)
±(3.3 mmHg +8 % rel.)
Anesthetic Gas Measurement (infrared spectrometry)
All values in Vol.% refer to ambient pressure 760 mmHg
Measuring range, anesthetic agent
Halothane 0 to 8.5 Vol.%
Isoflurane 0to8.5Vol.%
Enflurane 0 to 10 Vol.%
Sevoflurane 0to10Vol.%
Desflurane 0to20Vol.%
Resolution of the measurement 0.1 Vol.%
Resolution of the displayed value
(for insp. and exp. anesthetic agent)
0.1 Vol.%
Accuracy (at respiration rates of up to
60 bpm and I:E ratio of 1:1)
±(0.2 Vol.% + 15% rel.)