Specifications
Instructions for Use Apollo SW 4.5n 299
Part Number: 9053586, 3rd edition
Fresh-gas and Agent Consumption Measurement
Fresh-gas consumption per case 0 L to 9999 L per gas (O2, N2O, Air) (O2 value not including gas used for
the O2 flush and the auxiliary O2 flow meter)
Accuracy ±10% or ±1 L, whichever is greater
Resolution 1 L
Total agent consumption per case (liquid agent) 0 mL to 3000 mL per agent (Halothane, Isoflurane, Enflurane,
Sevoflurane, Desflurane)
Accuracy Typ. ±25% or ±2 mL, whichever is greater
Resolution 1 mL
Agent consumption due to patient uptake per
case (liquid agent)
0 mL to 3000 mL per agent (Halothane, Isoflurane, Enflurane,
Sevoflurane, Desflurane)
Accuracy Typ. ±25% or ±2 mL, whichever is greater
Resolution 1 mL
Soda lime consumption 0 to 1000 L (pure gas CO
2)
Accuracy Typ. ±30% or ±15 L, whichever is greater
Resolution of limit setting 10 L
1) Exception: If a measured desflurane concentration of at least 4Vol.% is present, a mixed agent identification is available as soon as the measured concentration
of the secondary agent reaches at least 10% of the desflurane concentration.
2) The respective value depends on the patient-gas measurement module (PGM) used which is displayed on System Information page.
SpO2 Measurement (optional) (light absorption)
Measuring range SpO2 1 % to 100%
Resolution of the displayed value 1%
Accuracy Depending on the sensor model, applies to DS-100 A.
Adults, within a range of 70 to 100% SpO
2 ±3%
Neonates, within a range of 70 to 100%
SpO2
±4%
Actualization time Once per pulse
Pulse rate 20 to 250 bpm
Resolution of the displayed value 1 bpm
Accuracy ±3 bpm