Specifications
Instructions for Use Apollo SW 4.5n 307
Part Number: 9053586, 3rd edition
Examples of subsequent changes to the network:
– Changing the network configuration
– Removing devices from the network
– Adding new devices to the network
– Performing upgrades or updates on devices that
are connected to the network
Information on connecting to the IT-net-
work
Prerequisites
This device must only be connected to the network
by service personnel. The IT representative of the
hospital must be consulted in advance.
The following documents must be observed:
– Accompanying documents of this device
– Descriptions of the network
– Description of the network-based alarm systems
Dräger recommends complying with IEC 80001-1
(risk management for IT-networks with medical
devices).
Serial ports
The following interfaces are supported:
– RS232 interfaces conforming to EIA RS-232
(CCITT V.24/V.28) for the following applications:
– MEDIBUS, MEDIBUS.X
– Connections with third party medical devices
Consequences of using an inappropriate network
If the network does not meet the requirements,
hazardous situations can result. The following
situations can occur with this device:
– Due to an insecure decentralized alarm system:
– Alarms are not transmitted.
– Alarms or data is transmitted with a delay.
– False alarms are triggered.
– During an interruption of the network connection:
– Alarms are not transmitted.
– Suppressed alarms or alarm tones are not
reactivated, but remain suppressed.