System Setup
Instructions for Use Apollo SW 4.5n 79
Part Number: 9053586, 3rd edition
CAUTION!
Risk of incorrect measured values
Silicone can enter the measuring cuvette and distort
the gas measurement.
Do not spray the O-rings of the water trap holder
with silicone spray.
CAUTION!
Risk of incorrect measured values
Aerosols can damage the diaphragm and the
measurement system.
Do not use aerosols in the breathing system. The
water trap must not be used in combination with a
medical nebulizer.

Table of Contents